Performance Characteristics

Agilent 54640-series Performance Characteristics

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Analog Channel Triggering

Range (Internal)

±6 div

Sensitivity*

<10 mV/div: Greater of 1 div or 5 mV

 

10 mV/div: 0.6 div

Coupling

ac (~10 Hz), dc, noise reject, HF reject and LF reject (~ 50 kHz)

 

Digital (D15 - D0) Channel Triggering (54641D and 54642D)

Threshold Range (used defined)

±8.0 V in 10 mV increments

Threshold Accuracy*

±(100 mV + 3% of threshold setting)

Predefined Thresholds

TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V

 

 

External (EXT) Triggering

 

Input Resistance

1 M±3% or 50-Ohm

Input Impedance

~ 13 pF

Maximum Input

CAT I 300 Vrms, 400 Vpk

 

CAT II 100 Vrms, 400 Vpk

 

with 10073C 10:1 probe:CAT I 500 Vpk, CAT II 400 Vpk

 

5 Vrms with 50-ohm input

Range

DC coupling: trigger level ±8 V

 

AC coupling/LF reject: AC input minus trigger level not to exceed ±8 V

Sensitivity

DC to 100 MHz, < 100 mV

 

100 MHz to max bandwidth, < 200 mV

Coupling

ac (~ 3.5 Hz), dc, noise reject, HF reject and LF reject (~ 50 kHz)

Probe Id (Agilent/HP &Tek

Auto probe sense for 54641A/42A

compatible)

 

7-18

Page 252
Image 252
Agilent Technologies 54621D, 24A, 42A, 54621A Analog Channel Triggering, Digital D15 D0 Channel Triggering 54641D and 54642D