Triggering the Oscilloscope

To set holdoff

To set holdoff

1Press the Mode/Coupling key.

2Turn the Entry knob to increase or decrease the trigger holdoff time shown in the Holdoff softkey.

Holdoff sets the amount of time that the oscilloscope waits before re-arming the trigger circuitry. Use Holdoff to stabilize the display of complex waveforms.

To get a stable trigger on the pulse burst shown below, set the holdoff time to be >200 ns but <600 ns.

+ROGRII

6FRSH￿WULJJHUV￿KHUH

￿￿￿￿QV

￿￿￿￿QV

By setting the Holdoff, you can synchronize triggers. The oscilloscope will trigger on one edge of the waveform, and ignore further edges until the holdoff time expires. The oscilloscope will then re-arm the trigger circuit to search for the next edge trigger. This allows the oscilloscope to trigger on a repeating pattern in a waveform.