Performance Characteristics

Agilent 54620-series Performance Characteristics

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Vertical System: Analog Channels

Analog channels

54621A/21D, 54622A/22D: Ch1 and 2 simultaneous acquisition

 

54624A: Ch 1, 2, 3, and 4 simultaneous acquisition

Bandwidth (-3dB)*

54621A/22D: dc to 60 MHz

 

54622A/22D/24A: dc to 100 MHz

ac coupled

54621A/21D: 3.5 Hz to 60 MHz

 

54622A/22D/24A: 3.5 Hz to 100 MHz

Calculated rise time

54621A/22D: ~5.8 ns

(= 0.35/bandwidth)

54622A/22D/24A: ~3.5 ns

Single Shot Bandwidth

50 MHz

Range1

1 mV/div to 5 V/div

Maximum Input

CAT I 300 Vrms, 400 Vpk

 

CAT II 100 Vrms, 400 Vpk

 

with 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk

Offset Range

±5 V on ranges <10 mV/div

 

±25 V on ranges 10 mV/div to 199 mV/div

 

±100 V on ranges 200 mV/div

Dynamic Range

Lesser of ±8 div or ±32 V

Input Resistance

1 M±1%

Input Capacitance

~ 14 pF

Coupling

ac, dc, ground

BW Limit

~ 20 MHz selectable

Channel-to-Channel Isolation

dc to 20 MHz > 40 dB

(with channels at same V/div)

20 MHz to max bandwidth > 30 dB

Probes

10:1 10074C shipped standard for each analog channel

Probe ID (Agilent/HP &

Auto probe sense

Tek Compatible)

 

11 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.

7-4

Page 238
Image 238
Agilent Technologies 22A, 54621D, 24A, 42A Agilent 54620-series Performance Characteristics, Vertical System Analog Channels