Making Measurements

Making Delay and Phase Measurements

Phase

Phase is the calculated phase shift from source 1 to source 2, expressed in degrees. Negative phase shift values indicate that the rising edge of source 1 occurred after the rising edge of source 2. Press the Settings softkey to select the source channels for the phase measurement.

Phase =

Delay

X 360

----------------------------------------Source 1 Period

Period

Source 1

Delay

Source 2

Press the Settings softkey to select the source 1 and source 2 channels for the phase measurement.

The default Phase settings measure from channel 1 to channel 2.

The example below shows a phase measurement between the channel 1 and the math d/dt function on channel 1.

Phase Measurement

5-44

Page 212
Image 212
Agilent Technologies 42A, 54621D, 24A, 54621A, 22A, 41A manual Phase Measurement