Triggering the Oscilloscope

To use Sequence triggering

To use Sequence triggering

Sequence trigger allows you to trigger the oscilloscope after finding a sequence of events. Defining a sequence trigger requires three steps:

aDefine an event to find before you search for the trigger.

The “find” event can be a pattern, an edge from a single channel, or the combination of a pattern and a channel edge.

bDefine the trigger event.

The “trigger on” event can be a pattern, an edge from a single channel, the combination of a pattern and a channel edge, or the nth occurrence of an edge from a single channel.

cSet an optional reset event.

If you choose to define a “reset” event, the event can be a pattern, an edge from a single channel, the combination of a pattern and a channel edge, or a timeout value.

Start

Yes

Yes

Find:?

Trigger on:?

No

No

No

Reset on:?

Scope triggers

Trigger holdoff

Yes

Sequence Trigger Flowchart

3-29