Agilent 54640-series Performance

Characteristics

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Acquisition: Analog Channels

Max Sample rate

2 GSa/s

Max Memory Depth

8 M

Vertical Resolution

8 bits

Peak Detection

1 ns @ max sample rate

Averages

selectable from 2, 4, 8, 16, 32, 64 ...to 16383

High Resolution Mode

12 bits of resolution when 100 us/div, average mode with average = 1

Filter:

Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope, whichever

 

is less) with vectors on.

Acquisition: Digital Channels (on 54641D and 54642D only)

Max Sample Rate

1 GSa/s

Max Memory Depth

8 MB

Vertical Resolution

1 bit

Glitch Detection (min pulse width)

5 ns

7-13

Page 247
Image 247
Agilent Technologies 24A, 54621D, 42A, 54621A, 22A, 41A manual Agilent 54640-series Performance Characteristics