Triggering the Oscilloscope

To use TV triggering

Triggering on Line 71

Line Numbers per Field for Each TV Standard

TV Standard

Field 1

Field 2

Alt Fld

NTSC

1 to 263

1 to 262

1 to 262

PAL

1 to 313

314 to 625

1 to 313

PAL-M

1 to 263

264 to 525

1 to 263

SECAM

1 to 313

314 to 625

1 to 313

GENERIC

1 to 1024

1 to 1024

1 to 1024 (Vertical)

Line Number Represents Count

In Generic mode, the line number represents the number of a count instead of a real line number. This is reflected in the label in the softkey changing from Line to Count. In the Mode softkey selections, Line:Field 1, Line:Field 2 and Count:Vertical are used to indicate where the counting starts. For an interlaced TV system, the counting starts from the rising edge of the first vertical serration pulse of Field 1 and/or Field 2. For a non-interlaced TV system, the counting starts after the rising edge of the vertical sync pulse.