Performance Characteristics

Agilent 54640-series Performance Characteristics

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Horizontal

Range

1 ns/div to 50 s/div

Resolution

2.5 ps

Vernier

1-2-5 increments when off, 25 minor increments between major settings when on

Reference Positions

Left, Center, Right

Delay Range

 

Pre-trigger (negative delay)

Greater of 1 screen width or 1 ms

Post-trigger (positive delay)

500 seconds

Analog Delta-t Accuracy

 

Same Channel*

±0.005% reading ±0.2% screen width ±20 ps

 

Example: for signal with pulse width of 10 us, scope set to 5 us/div (50 us screen width),

 

delta-t accuracy = ±{.005%(10 us) + 0.2% (50 us) + 20 ps} = 100.52 ns

Channel-to-Channel

±0.005% reading ±0.2% screen width ±40 ps

Digital Delta-t Accuracy

(non-Vernier settings)

Same Channel

±0.005% reading ±0.2% screen width ±(1 digital sample period, 1 ns)

 

Example: for signal with pulse width of 10 us, scope set to 5 us/div (50 us screen width),

 

and single pod active (1 GSa/s), delta-t accuracy = ±{.005%(10 us) + 0.2% (50 us) + 1 ns}

 

= 101.5 ns

Channel-to-Channel

±0.005% reading ±0.2% screen width ±(1 digital sample period) ±(chan-to-chan skew))

Delay Jitter

10 ppm

RMS Jitter

0.025% screen width + 30 ps

Modes

Main, Delayed, Roll, XY

XY

 

Bandwidth

Max bandwidth

Phase error @ 1 MHz

1.8 degrees