5-45
Making Measurements
Making voltage measurements automatically
Making voltage measurements automatically
Measurement units for each input channel can be s et to Volts or Amps using
the channel Probe Units softk ey. A s cal e uni t of U (undefined) will be displayed
for math function 1-2 and for d/dt, and dt whe n 1- 2 or 1 +2 is t he s el e ct e d s o ur c e
if channel 1 and channel 2 are set to dissimilar units in the channel Probe Units
softkey.
The following figure shows the voltage measu rement points.
Math Measurements and Units
Only Peak-Peak, Maximum, Minimum, Average, X at Min, and X at Max automatic
measurements may be made on a math FFT function. See Making time
measurements automatically for the FFT X at Max and X at Min measurement. Use
the cursors to make other measurements on F FT. All voltage measurements ma y be
made on other math functions. The resultant units are:
FFT:
1*2:
12:
d/dt:
dt:
dB* (decibels)
V2, A2 or W (Volt-Amp)
V (Volts) or A (Amps)
V/s or A/s (V/second A/second)
Vs or As (V-seconds or A-seconds)
* When the FFT source is channel 1 or channel 2 (or channel 3 or 4 on the 54624A), FFT unit s
will be displayed in dBV when channel units is set to Volts and chan nel impedance is set to
1M
. FFT units will be displayed in dBm when channel units is set to Volts and chann el
impedance is set to 50 (54640-series only). FFT units will be displaye d as dB for all other FFT
sources or when a source channels units has been set to Amps.