Making Measurements

Making voltage measurements automatically

Making voltage measurements automatically

Measurement units for each input channel can be set to Volts or Amps using the channel Probe Units softkey. A scale unit of U (undefined) will be displayed for math function 1-2 and for d/dt, and dt when 1-2 or 1+2 is the selected source if channel 1 and channel 2 are set to dissimilar units in the channel Probe Units softkey.

Math Measurements and Units

Only Peak-Peak, Maximum, Minimum, Average, X at Min, and X at Max automatic measurements may be made on a math FFT function. See “Making time measurements automatically” for the FFT X at Max and X at Min measurement. Use the cursors to make other measurements on FFT. All voltage measurements may be made on other math functions. The resultant units are:

FFT:

dB* (decibels)

1

* 2:

V2, A2 or W (Volt-Amp)

1

– 2:

V (Volts) or A (Amps)

d/dt:

V/s or A/s (V/second A/second)

dt:

Vs or As (V-seconds or A-seconds)

*When the FFT source is channel 1 or channel 2 (or channel 3 or 4 on the 54624A), FFT units

will be displayed in dBV when channel units is set to Volts and channel impedance is set to 1 M. FFT units will be displayed in dBm when channel units is set to Volts and channel impedance is set to 50(54640-series only). FFT units will be displayed as dB for all other FFT sources or when a source channel’s units has been set to Amps.

The following figure shows the voltage measurement points.

Maximum

Top

Amplitude

Peak-Peak

 

Base

 

Minimum

5-45

Page 213
Image 213
Agilent Technologies 54621A, 54621D, 24A, 42A, 22A, 41A Making voltage measurements automatically, Math Measurements and Units