Making Measurements

To reduce the random noise on a signal

Low-frequency reject (LF reject) adds a high-pass filter with the 3-dB point at 50 kHz. Use LF reject to remove low-frequency signals, such as power line noise, from the trigger path.

0 dB

3 dB down point

Pass

Band

dc 50 kHz

LF Reject

Noise reject increases the trigger hysteresis band. By increasing the trigger hysteresis band, you reduce the possibility of triggering on noise. However, this also decreases the trigger sensitivity so that a slightly larger signal is required to trigger the oscilloscope.

Random noise on the displayed waveform

5-7

Page 175
Image 175
Agilent Technologies 24A, 54621D, 42A, 54621A, 22A, 41A manual LF Reject