Front-Panel Overview

To make automatic measurements

To make automatic measurements

You can use automatic measurements on any channel source or any running math function. Cursors are turned on to focus on the most recently selected measurement (right-most on the measurement line above the softkeys on the display).

1 Press the Quick Meas key to display the automatic measurement menu.

2Press the Source softkey to select the channel or running math function on which the quick measurements will be made.

Only channels or math functions that are displayed are available for measurements. If you choose an invalid source channel for a measurement, the measurement will default to the nearest in the list that makes the source valid.

If a portion of the waveform required for a measurement is not displayed or does not display enough resolution to make the measurement, the result will be displayed with a message such as greater than a value, less than a value, not enough edges, not enough amplitude, incomplete, or waveform is clipped to indicate that the measurement may not be reliable.

3Press the Clear Meas softkey to stop making measurements and to erase the measurement results from the measurement line above the softkeys.

When Quick Meas is pressed again, the default measurements on an analog channel will be will be Frequency and Peak-Peak.

4 Choose what measurement you want on that source by pressing the

Select softkey, then turn the Entry knob to select the desired measurement from the popup list.

5 Press the Measure softkey to make the selected measurement.

6To turn off Quick Meas, press the Quick Meas key again until it is not illuminated.

For detailed information about making automatic measurements, refer to the “Making Measurements” chapter.

2-23

Page 61
Image 61
Agilent Technologies 24A, 42A To make automatic measurements, Choose what measurement you want on that source by pressing