C A U T I O N

C A U T I O N

Getting Started

To connect the oscilloscope analog probes

To connect the oscilloscope analog probes

The analog input impedance of these oscilloscopes is selectable either 50(54640-series only) or 1 M. The 50mode matches 50cables commonly used in making high frequency measurements. This impedance matching gives you the most accurate measurements since reflections are minimized along the signal path. The 1 Mmode is for use with probes and for general purpose measurements. The higher impedance minimizes the loading effect of the oscilloscope on the circuit under test.

Do not exceed 5 Vrms in 50mode on the 54640-series models. Input protection is enabled in 50mode and the 50load will disconnect if greater than 5 Vrms is detected. However the inputs could still be damaged, depending on the time constant of the signal.

The 50input protection mode on the 54640-series models only functions when the oscilloscope is powered on.

1Connect the supplied 1.5-meter, 10:1 oscilloscope probe to an analog channel BNC connector input on the oscilloscope.

Maximum input voltage for analog inputs:

CAT I 300 Vrms, 400 Vpk

CAT II 100 Vrms, 400 Vpk

with 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk

2Connect the retractable hook tip on the probe tip to the circuit point of interest. Be sure to connect the probe ground lead to a ground point on the circuit.

The probe ground lead is connected to the oscilloscope chassis and the ground wire in the power cord. If you need to connect the ground lead to a point in the circuit that cannot be grounded to power ground, consider using a differential probe.

1-15

Page 25
Image 25
Agilent Technologies 24A, 54621D, 42A, 54621A, 22A, 41A manual To connect the oscilloscope analog probes