Triggering the Oscilloscope

To use Duration triggering

To use Duration triggering

Duration trigger lets you define a pattern, then trigger on a specified time duration of this logical AND combination of the channels.

1Press the More key in the Trigger section of the front panel, rotate the Entry knob until Duration is displayed in the Trigger softkey, then press the Settings softkey to display the Duration trigger menu.

Duration

Selected

Trigger level

trigger

channel

or threshold

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Channel

Logic level

Qualifier

Qualifier time set

Return to

select

 

 

 

 

 

previous menu

2For each analog or digital channel you want to include in the desired pattern, press the Channel softkey to select the channel.

This is the channel source for the H, L, or X condition. As you press the Channel softkey (or rotate the Entry knob on mixed-signal oscilloscopes), the channel you select is highlighted in the Pattern = line directly above the softkeys and in the upper-right corner of the display next to "Dur". External trigger may also be specified as a channel in the pattern when using the 2-channel and 4-channel oscilloscopes.

Adjust the trigger level for the selected analog channel by turning the Trigger Level knob. The position of the trigger level for the analog channel is indicated by the trigger level icon T at the far left side of the display when DC coupling is selected Set the threshold for the selected digital channel group by pressing the D7 Thru D0 or D15 Thru D8 key then press the Threshold softkey. The value of the trigger level or digital threshold is displayed in the upper-right corner of the display.

3-21

Page 83
Image 83
Agilent Technologies 41A, 54621D, 24A, 42A, 54621A, 22A manual To use Duration triggering