Performance Characteristics

Agilent 54620-series Performance Characteristics

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Horizontal

Range

5 ns/div to 50 s/div

Resolution

25 ps

Vernier

1-2-5 increments when off, 25 minor increments between major settings when on

Reference Positions

Left, Center, Right

Delay Range

 

Pre-trigger (negative delay)

Greater of 1 screen width or 10 ms

Post-trigger (positive delay)

500 seconds

Analog Delta-t Accuracy

 

Same Channel*

±0.01% reading ±0.2% screen width ±40 ps

 

Example: for signal with pulse width of 10 us, scope set to 5 us/div (50 us screen width),

 

delta-t accuracy = ±{.01%(10 us) + 0.2% (50 us) + 40 ps} = 101.04 ns

Channel-to-Channel

±0.01% reading ±0.2% screen width ±80 ps

Digital Delta-t Accuracy

(non-Vernier settings)

Same Channel

±0.01% reading ±0.2% screen width ±(1 digital sample period, 2.5 or 5 ns based on

 

sample rate of 200/400 MSa/s)

 

Example: for signal with pulse width of 10 us, scope set to 5 us/div (50 us screen width),

 

and single pod active (400 MSa/s), delta-t accuracy = ±{.01%(10 us) + 0.2% (50 us) + 2.5

 

ns} = 103.5 ns

Channel-to-Channel

±0.01% reading ±0.2% screen width ±(1 digital sample period, 2.5 or 5 ns)

 

±chan-to-chan skew (2 ns typical, 3 ns maximum)

Delay Jitter

10 ppm

RMS Jitter

0.025% screen width + 30 ps

Modes

Main, Delayed, Roll, XY

XY

 

Z blanking

1.4 V blanks trace (use External trigger)

Bandwidth

Max bandwidth

Phase error @ 1 MHz

1.8 degrees

7-6

Page 240
Image 240
Agilent Technologies 54621D, 24A, 42A, 54621A, 22A, 41A manual Horizontal, Sample rate of 200/400 MSa/s