Triggering the Oscilloscope

To use Sequence triggering

5If you select an edge term, one channel must be set to a rising edge or a falling edge. All other channel edges will be set to don’t care (X).

a Press the Channel softkey (or rotate the Entry knob on mixed-signal oscilloscopes) to select the channel.

As you select a channel, the channel is highlighted in the selected pattern list shown in the waveform area.

bThen press the X softkey to select rising edge or falling edge. All other channels will be defaulted to don’t care (X).

If you want to reassign an edge to a different channel, repeat the above step. The value of the original channel edge will be defaulted to X (don’t care).

Define the Trigger on: stage

1Press the Stage softkey and select Trigger on:.

Trigger on: is the next stage in the trigger sequence. When you select the Stage

Trigger on: softkey, the next softkey to the right will display Trigger: and give you

a list of the terms you can define for the Trigger on: stage. The Trigger on: stage can be set to one of the following conditions:

Pattern 2 Entered – A pattern is entered on the last edge that makes the pattern true (logical AND).

Pattern 2 Exited – A pattern is exited on the first edge that makes the pattern false (logical NAND).

Edge 2

Pattern 2 and Edge 2

Nth Edge 2

2 Press the Trigger: softkey and select a stage to trigger on.

3To define the term(s) used in the Trigger on: stage, press the Term softkey and select the pattern or edge term displayed in the Trigger: softkey.

4If you select a pattern term, each channel in the pattern must be set to a H (high), L (low), or X (don’t care).

a Press the Channel softkey (or rotate the Entry knob on mixed-signal oscilloscopes) to select the channel.

b Press the L H X softkey to set a level for the channel. c Repeat for all channels in the pattern.

3-32

Page 94
Image 94
Agilent Technologies 22A, 54621D, 24A, 42A Press the Stage softkey and select Trigger on, Edge Pattern 2 and Edge Nth Edge