Triggering the Oscilloscope

To use SPI triggering

Set up the number of bits in the serial data string and set values for those data bits

1Turn the Entry knob to set the number of bits (#Bits) in the serial data string. The number of bits in the string can be set anywhere from 4 bits to 32 bits. The data values for the serial string are displayed in the Data string in the waveform area.

2Turn the Entry knob to select a specific data bit in the serial string as shown in the Bit softkey.

As you rotate the Entry knob, the bit is highlighted in the Data string shown in the waveform area.

3Press the L H X H (high), or X

softkey to set the bit selected in the Bit softkey to L (low), (don’t care).

Repeat steps 2 and 3 to assign values for all bits.

Resetting all bits in the serial data string to one value.

To reset all of the bit values in the serial data string to one value (L, H, or X):

a Press the L H X softkey to select your reset bit value.

b Press the Set all Bits softkey to reset the data string to the selected value.

3-36

Page 98
Image 98