Performance Characteristics

Agilent 54620-series Performance Characteristics

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Vertical System: Analog Channels (continued)

ESD Tolerance

±2 kV

Noise Peak-to-Peak

2% full scale or 1 mV, whichever is greater

Common Mode Rejection Ratio

20 dB @ 50 MHz

DC Vertical Gain Accuracy*1

±2.0% full scale

DC Vertical Offset Accuracy Value

< 200 mV/div: ±0.1 div ±1.0 mV ±0.5% offset

 

200 mV/div: ±0.1 div ±1.0 mV ±1.5% offset value

Single Cursor Accuracy1

±{DC Vertical Gain Accuracy + DC Vertical Offset

 

Accuracy + 0.2% full scale (~1/2 LSB) }

 

Example: For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,

 

accuracy = ±{2.0%(80mV) + 0.1 (10 mV) + 1.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} = ± 3.78 mV

Dual Cursor Accuracy*1

±{DC Vertical Gain Accuracy + 0.4% full scale (~1 LSB)}

 

Example: For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,

 

accuracy = ±{2.0%(80 mV) + 0.4%(80 mV)} = ±1.92 mV

11 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.

Vertical System: Digital Channels (54621D and 54622D only)

Number of Channels

16 Digital – labeled D15 – D0

Threshold Groupings

Pod 1: D7 - D0

 

Pod 2: D15 - D8

Threshold Selections

TTL, CMOS, ECL, user-definable (selectable by pod)

User-Defined Threshold Range

±8.0 V in 10 mV increments

Maximum Input

±40 V peak CAT I

Voltage

 

Threshold Accuracy*

±(100 mV + 3% of threshold setting)

Input Dynamic Range

±10 V about threshold

Minimum Input Voltage Swing

500 mV peak-to-peak

Input Capacitance

~ 8 pF

Input Resistance

100 k±2% at probe tip

Channel-to-Channel Skew

2 ns typical, 3 ns maximum

7-5

Page 239
Image 239
Agilent Technologies 41A, 24A, 42A, 54621A, 22A manual Vertical System Digital Channels 54621D and 54622D only