Triggering the Oscilloscope

To use USB triggering

To use USB triggering

USB trigger will trigger on a Start of Packet (SOP), End of Packet (EOP) signal, Reset Complete (RC), Enter Suspend (Suspend), or Exit Suspend (Exit Sus) on the differential USB data lines (D+ and D-). USB Low Speed and Full Speed are supported by this trigger.

End of packet trigger

D+,–

Bus Idle

D–,+

SOP

SE0 portion of EOP

1Press the More key in the Trigger section of the front panel, rotate the Entry knob until USB is displayed in the Trigger softkey, then press the Settings softkey to display USB trigger menu.

Currently selected

Trigger level

data source

or threshold

USB trigger

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Signal

D+ source

D– source

Trigger on:

Return to

speed

channel

channel

condition

previous menu

2Press the Speed softkey to select the speed of the transaction being probed.

You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s).

3-50

Page 112
Image 112
Agilent Technologies 22A, 54621D, 24A, 42A, 54621A, 41A manual To use USB triggering