Making Measurements

To use the XY horizontal mode

6Move the Y1 and Y2 cursors to the intersection of the signal and the Y axis.

Again, note the Y value.

Cursors set to center of signal

7 Calculate the phase difference using the formula below.

sin θ =

second--------------------------Y

=

0.648------------

; θ = 32.68 degrees of phase shift

 

first Y

 

1.20

 

XY Display Mode Operating Hint (Z-axis input on 54620-series only)

When you select the XY display mode, the time base is turned off. Channel 1 is the X-axis input, channel 2 is the Y-axis input, and the External trigger is the Z-axis input. If you only want to see portions of the Y versus X display, use the Z-axis input. Z-axis turns the trace on and off (analog oscilloscopes called this Z-axis blanking because it turned the beam on and off). When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V), the trace is turned off.

5-15

Page 183