MegaZoom Concepts and Oscilloscope Operation

To run and stop an acquisition

To run and stop an acquisition

To begin an acquisition, press Run/Stop (key is illuminated green when running).

The instrument begins acquiring data while searching for a trigger condition. If a trigger occurs, the acquired data is shown in the display.

To stop an acquisition in process, press Run/Stop (key is illuminated red when stopped).

The instrument stops acquiring data and STOP is displayed in the trigger mode position of the status line in the upper-right corner of the display. If the instrument was triggered (even by auto triggering) and the pre-trigger and post-trigger buffers are full, the results are displayed on the screen. If the buffers are not full (that is, no triggers occurred), the waveform display area will show nothing.

To take a single trace

Press the Single key.

The Single key will illuminate. The oscilloscope starts the acquisition system and begins searching for the trigger condition.

When the oscilloscope is in infinite persistence display mode (see “To use Infinite Persistence” topic) pressing Single erases the results of previous acquisitions from the current display contents.

Auto Single

Auto-trigger will generate a trigger for you if one is not found in the predetermined time from when the trigger system is armed. To take a single-shot acquisition, and you are not particularly interested in triggering the acquisition (for example, if you are probing a DC level), set Trigger Mode/Coupling mode to Auto and press the Single key. If a trigger exists, it will be used; if a trigger does not exist, an untriggered acquisition will be taken.

4-28

Page 142
Image 142
Agilent Technologies 22A, 54621D To run and stop an acquisition, To take a single trace, Press the Single key, Auto Single