Making Measurements

Making overshoot and preshoot measurements

Making overshoot and preshoot measurements

Digital channel time measurements

Automatic measurements Preshoot and Overshoot are not valid measurements for the math FFT function or for digital channels on the mixed-signal oscilloscope.

Measurement definitions

Preshoot

Preshoot is distortion that precedes a major edge transition expressed as a percentage of Amplitude. The X cursors show which edge is being measured (edge closest to the trigger reference point).

Base – local Minimum

X 100

Rising edge preshoot = ------------------------------------------------------

Amplitude

 

local Maximum – Top

X 100

Falling edge preshoot = -----------------------------------------------------

Amplitude

 

local Maximum

Preshoot

 

Top

Base

Preshoot

local Minimum

5-48