Getting Started

To use the digital probes (mixed-signal oscilloscope only)

To use the digital probes (mixed-signal oscilloscope only)

1If you feel it’s necessary, turn off the power supply to the circuit under test.

Off

Turning off power to the circuit under test would only prevent damage that might occur if you accidentally short two lines together while connecting probes. You can leave the oscilloscope powered on because no voltage appears at the probes.

2Connect the digital probe cable to D15 - D0 connector on the front panel of the mixed-signal oscilloscope. The digital probe cable is indexed so you can connect it only one way. You do not need to power-off the oscilloscope.

Use only the Agilent part number 54620-68701 digital probe kit supplied with the mixed-signal oscilloscope. Additional probe kits may be ordered by specifying Agilent part number 10089A.

1-17

Page 27
Image 27
Agilent Technologies 54621A, 54621D, 24A, 42A, 22A, 41A manual To use the digital probes mixed-signal oscilloscope only