Performance Characteristics

Agilent 54640-series Performance Characteristics

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Vertical System: Analog Channels

Analog channels

Ch1 and 2 simultaneous acquisition

Bandwidth (-3dB)*

54641A/42D: dc to 350 MHz

 

54642A/42D: dc to 500 MHz

ac coupled

54641A/41D: 3.5 Hz to 350 MHz

 

54642A/42D: 3.5 Hz to 500 MHz

Calculated rise time

54641A/42D: ~1.0 ns

(= 0.35/bandwidth)

54642A/42D: ~700 ps

Single Shot Bandwidth

54641A/41A: 350 MHz maximum

 

54642A/42D: 500 MHz maximum

Range1

2 mV/div to 5 V/div

Maximum Input

CAT I 300 Vrms, 400 Vpk

 

CAT II 100 Vrms, 400 Vpk

 

with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk

 

5 Vrms with 50-ohm input

Offset Range

±5 V on ranges <10 mV/div

 

±20 V on ranges 10 mV/div to 200 mV/div

 

±75 V on ranges >200 mV/div

Dynamic Range

Lesser of ±8 div or ±32 V

Input Resistance

1 M±1%, 50-ohm selectable

Input Capacitance

~ 13 pF

Coupling

ac, dc

BW Limit

~ 25 MHz selectable

Channel-to-Channel Isolation

DC to max bandwidth > 40 dB

(with channels at same V/div)

 

Probes

10:1 10073C shipped standard for each analog channel

Probe ID (Agilent/HP &

Auto probe sense

Tek Compatible)

 

12 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.

7-14

Page 248
Image 248
Agilent Technologies 42A, 54621D, 24A, 22A Agilent 54640-series Performance Characteristics, Vertical System Analog Channels