Getting Started

To compensate your analog probe

To compensate your analog probe

You should compensate you analog probes to match their characteristics to the oscilloscope. A poorly compensated probe can introduce measurement errors. To compensate a probe, follow these steps:

1Connect the probe from channel 1 to the Probe Comp signal on the lower- right corner of the front panel.

2Press Autoscale.

3Use a nonmetallic tool to adjust the trimmer capacitor on the probe for the flattest pulse possible.

Perfectly compensated

Over compensated

Under compensated

comp.cdr