29Transition Parameter Setting
Users could press
Shift
+
S-Tran
to set the transition parameter.
Shift
+
S-Tran
LEVEL A = *****
Setup value A
Enter
WIDTH A = ******
Setup time width of valu e A
Enter
LEVEL B=*****
Setup value B
Enter
WIDTH B= *****
Setup time width of valu e B
Enter
CONTINUOUS
PULSE
TOGGLED
Choose one of the three transition
modes
Enter
Finish transition s etting
Continuous Transient Operation In this mode, electronic load will generates a repetitive pulse stream that toggles between two load levels. ; Load could switch the st ate between two sett ing value (value A and v alue B).
In this following example, assume that the CC mode is active; the applicable transient operation parameters have
been set as follows.
For example:
Continuous mode, current level A =5A, width = 2mS. Current level B =10A, width = 3ms. Testing machine input
voltage is 12V.
10A5AContinuous Transient Operation2.0ms 3.0msAction
1. Select the operation mode (
CC,CV,CR or CP), and set a proper value, press
On/Off
to turn off the load input.
2. Press
Shift
+
S-Tran, set LEVEL A=5A, press
Enter
, set WIDTH A=3mS, press
Enter
, set
LEVER B=10A, press
Enter
, set WIDTH B=2mS, press
Enter
to confirm.
3.Transition mode now is
CONTINOUS, press
Enter
to confirm.
4. Press
Shift
+ Tranto activate the transient mode.
5. Press
Shift
+ Tran
again to stop the transient operation.
6. Press
On/Off
, the input of load is powered on.