Transition Parameter Setting
Users could press Shift | + | |||
| Shift + |
| LEVEL A = ***** | Setup value A |
| Enter |
| WIDTH A = ****** | Setup time width of value A |
| Enter |
| LEVEL B=***** | Setup value B |
| Enter |
| WIDTH B= ***** | Setup time width of value B |
| Enter |
| CONTINUOUS | Choose one of the three transition |
|
|
| PULSE | modes |
|
|
| TOGGLED |
|
| Enter |
|
| Finish transition setting |
|
|
|
|
|
Continuous Transient Operation
In this mode, electronic load will generates a repetitive pulse stream that toggles between two load levels. ; Load could switch the state between two setting value (value A and value B).
In this following example, assume that the CC mode is active; the applicable transient operation parameters have been set as follows.
For example:
Continuous mode, current level A =5A, width = 2mS. Current level B =10A, width = 3ms. Testing machine input voltage is 12V.
10A |
|
5A |
|
2.0ms | 3.0ms |
Continuous Transient Operation |
Action
1. | Select the operation mode (CC,CV,CR or CP), and set a proper value, press On/Off to turn off the load input. | ||||||
2. | Press | Shift | + | , set LEVEL A=5A, press Enter | , set | WIDTH A=3mS, press Enter , set | |
| LEVER B=10A, | press Enter , set WIDTH B=2mS, press | Enter | to confirm. | |||
3.Transition mode now is CONTINOUS, press Enter to confirm. |
| ||||||
4. | Press | Shift | + | Tran | to activate the transient mode. |
|
|
5. | Press | Shift | + | Tran | again to stop the transient operation. |
|
6. Press On/Off , the input of load is powered on.
Pulse Transient Operation
In this mode, generates a transient pulse of programmable width when pulsed transient operation is in effect.
29