32
18) Press
Enter
to confirm.

Quick Recall Function
This function allows you to recall 10 groups numbers which were stored previously.
Steps:
1) Press
Shift
+
Menu
2) Press
to
CONFIG
, press
Enter
into submenu, press
to
SHORTCURT RECALL
3) Press
Enter
, select
ON
and press
Enter
to confirm.
4) Press
ESC
two times and quit the menu.
5) Press
1
and y ou can get the first gr oup number. Pres s
2
, you can get the second group number. Pres s
0
,
you can get the tenth group number. And if there is not stored number, VFD will show
NO EEPROM DATA
.

Automatic Testing Function of Electronic Load

The CD load’s firmware provides an automatic testing function that is convenient for testing various

parameters of power supplies with different load conditions. Now we take a kind of charger as an

example to explain the usage of the automatic testing of the electronic load.

The test proceeding of R35 charger
Work produce process method
Mode Voltage(V) Current(mA) Power(W) Ripple
counter(mVpp)
The first step CC 5.8
6.15 210 4(input175VAC)
The second step None 5.9
6.4 0 1.2(input175VAC)
The third step CC 5 205
245
The fourth step CV 3 205
245
The fifth step CV 2 205
245
The sixth
step
Short
circuit
0 245
50mVpp

Edit the Testing File

1) Press
Shift
+
Menu
into menu, VFD displays
CONFIG
2) Press
to
LIST SET
.
3) Press
Enter
into s ubmenu, press
to
EDIT LIST FILE
.