18) Press

Enter

to confirm.

 

Quick Recall Function

This function allows you to recall 10 groups numbers which were stored previously.

Steps:

1) Press Shift + Menu

2)

Press▼to CONFIG, press Enter into submenu, press▼ to SHORTCURT RECALL

3)

Press Enter , select ON and press Enter to confirm.

4)

Press ESC

two times and quit the menu.

5)

Press 1

and you can get the first group number. Press 2 , you can get the second group number. Press 0 ,

you can get the tenth group number. And if there is not stored number, VFD will show NO EEPROM DATA.

Automatic Testing Function of Electronic Load

The CD load’s firmware provides an automatic testing function that is convenient for testing various parameters of power supplies with different load conditions. Now we take a kind of charger as an example to explain the usage of the automatic testing of the electronic load.

The test proceeding of R35 charger

Work produce

process

 

 

 

 

method

 

 

 

 

 

 

 

 

 

 

Mode

Voltage(V)

Current(mA)

Power(W)

Ripple

 

 

 

 

 

 

 

counter(mVpp)

 

 

 

 

 

 

 

 

The first step

CC

5.86.15

210

≤4(input175VAC)

 

 

 

 

 

 

 

 

 

 

The second step

None

5.96.4

 

0

≤1.2(input175VAC)

 

 

 

 

 

 

 

 

 

The third step

CC

5

205245

 

 

 

 

 

 

 

 

 

≤50mVpp

 

The fourth step

CV

3

205

 

 

245

 

 

 

The fifth step

CV

2

205245

 

 

 

 

 

 

 

 

 

 

The sixth step

Short

0

≤245

 

 

 

 

circuit

 

 

 

 

 

 

 

 

 

 

 

 

 

Edit the Testing File

1) Press Shift + Menu into menu, VFD displays CONFIG

2)Press▼ to LIST SET.

3) Press Enter into submenu, press▼ toEDIT LIST FILE.

32

Page 32
Image 32