30
For example: When load receiving one trigger signal, it will switch to 10A current value, and taking 10mS to return the
current value of 5A.
10A
5A
Pulsed Transient Operation
10ms
TWD 10ms
TWD
TRIG TRIG

Action

1. Select the operation mode (
CC,CV,CR or CP
), and set a proper value, press
On/Off
to turn off the load input.
2. Press
Shift
+
S-Tran
,
set LEVER A=5A, LEVER B=10A, WIDTH B=10mS. Transition mode now is
CONTINOUS
.
3. Press
till >PULSE
.
4. Press
Enter
, set transient mode is
PULSE
.
5. Press
Shift
+
Tran
to activate the transient mode.
6. Press
Shift
+
Trigger
to start another pulse. Press
Shift
+
Trigger
, get more pulse.
7. Press
Shift
+
Tran
again to stop the transient operation.
8. Press
On/Off
,
the input of load is powered on.

.

Toggled Transient Operation
In this mode, after transition operation start, Load could change the input between the main level and the
transient level when toggled transient operation is in effect.

For example:
When Load receives one trigger signal, Load current will switch between 5A and 10A.
10A
5A
TRG TRG

Toggled Transient Operation

Steps

1. Press
Shift
+
S-Tran
, Setting LEVER A=5A, LEVER B=10A, transition mode is TOGGLED.
2. Press
Shift
+
Tran
to activate the transient mode.