For example: When load receiving one trigger signal, it will switch to 10A current value, and taking 10mS to return the current value of 5A.

10A

 

 

5A

TWD

TWD

 

10ms

10ms

 

TRIG

TRIG

 

Pulsed Transient Operation

Action

1.

Select the operation mode (CC,CV,CR or CP), and set a proper value, press

On/Off to turn off the load input.

2. Press

Shift

+

S-Tran

, set LEVER A=5A, LEVER B=10A, WIDTH B=10mS. Transition mode now is

CONTINOUS.

 

 

 

 

3.

Press ▼ till >PULSE.

 

 

4.

Press

Enter

, set transient mode is PULSE.

 

5.

Press

Shift

+

Tran

to activate the transient mode.

 

6.

Press

Shift

+ Trigger

to start another pulse. Press Shift + Trigger

, get more pulse.

7.

Press

Shift

+

Tran

again to stop the transient operation.

 

8. Press On/Off , the input of load is powered on.

.

Toggled Transient Operation

In this mode, after transition operation start, Load could change the input between the main level and the transient level when toggled transient operation is in effect.

For example:

When Load receives one trigger signal, Load current will switch between 5A and 10A.

10A

 

5A

 

TRG

TRG

Toggled Transient Operation

Steps

1. Press

Shift

+ S-Tran , Setting LEVER A=5A, LEVER B=10A, transition mode is TOGGLED.

2. Press

Shift

+ Tran to activate the transient mode.

30

Page 30
Image 30
B&K 8510 instruction manual Toggled Transient Operation, Steps