CW Mode Regulation

0-100W

 

±(1%+0.1%FS)

 

 

1mW

 

Input current ≥FS 10%

 

 

 

 

 

 

100-600W

 

±(1%+0.1%FS)

 

 

100mW

Input Voltage≥FS 10%

 

 

 

 

 

 

 

 

 

 

 

Current Measurement

0-12A

 

±(0.1% + 0.1%FS)

 

1mA

 

 

 

 

 

 

 

 

0-120A

 

±(0.2%+0.15%FS)

 

10mA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Voltage Measurement

 

 

0-18V

 

±(0.02% + 0.02%FS)

 

1mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0-120V

 

±(0.02% + 0.025%FS)

 

10mV

 

 

 

 

 

 

 

 

 

Power Measurement

0-100W

 

±(1%+0.1%FS)

 

 

1mW

 

 

Input current≥FS 10%

 

 

 

 

 

 

 

 

 

 

 

 

 

100-600W

 

±(1%+0.1%FS)

 

 

100mW

Input Voltage≥FS 10%

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Input=0.8-120V

Max measurement capacity= 999A/H

 

Battery testing function

 

 

 

 

 

 

Resolution =10mA

 

Timer range=1~60000sec

 

 

 

 

 

 

 

 

 

 

Transition Mode

Range o Frequency 0.1Hz-1kHz

 

Frequency error rate0.5%

 

 

Weight

 

 

30.8 lb. (14kg)

 

 

 

 

 

 

 

 

 

 

 

NOTE: Specifications and information are subject to change without notice. Please visit www.bkprecision.com for the most current product information.

Remote Operation Mode

The DB9 connector in the rear panel of the electronic load can be connected to a RS-232 or USB interface via a communication cable. The following information may help you to know how to control the input of electronic load through PC.

1 Communication cable

IT-E131 RS232 Communication cable

The DB9 interface connector on the rear panel of electronic load is TTL voltage level; you can use the communication cable (IT-E131) to connect the DB9 interface connector of the electronic load and the RS-232 interface connector of computer for the communication.

 

IT-E131 communication

 

PC

cable

Load

 

COMPUTER

IT IT-E131 ISOLATED COMMUNICATION CABLE

RS232 ISOLATION TTL(5V)

859666668889942311

RX

TX

INSTRUMENT

IT-E132 USB Communication cable

36

Page 36
Image 36
B&K 8510 instruction manual Remote Operation Mode, IT-E131 RS232 Communication cable, IT-E132 USB Communication cable