15)Repeat 9)14), and set the other process step by step as follows:

A.CONST CURRENT,0A,SHORT OFF, READ BACK V, 5.9V, 6.4V,1S

B.CONST VOLTAGE, 5V, SHORT OFF,READ BACK A, 0.205A, 0.245A,1S

C.CONST VOLTAGE, 3V, SHORT OFF, READ BACK A, 0.205A,0.245A,1S

D.CONST VOLTAGE, 2V,SHORT OFF, READ BACK A, 0.205A,0.245A, 1S

E.CONST CURRENT, 0A, SHORT ON, READ BACK A,0A,0.245A, 1S

16)VFD displays SHORT TEST FILE*, which requests saving the files edited to EEPROM, automatically testing files and LIST files use a common memory area, and the area can contain 8 groups of automatic testing files for

next usage. Here, the testing file could be saved in the first group, press 1 and press Enter to confirm.

17) When the edition of automatic testing file is finished, press Enter two times to back the menu.

Taking out the Testing File Rapidly

1)

Press

Shift

+ Menu to enter into menu, VFD displays>CONFIG

2) Press▼ and move the menu to>LIST SET

3) Press

Enter

to enter into submenu, VFD displays>MODE SET

4)

Press▼ and move the menu to>CALL TEST FILE

5)

Press

Enter

and recall the file edited

Automatic Testing

After editing the automatic testing file, the automatic testing could be carried out by the below steps.

1) Press

Shift

+ I-set

to enter into automatic testing, VFD displays the name of current file: NAME: TEST

FILE1

 

 

 

2) Press

Enter

to observe the voltage and current we inputted, and also can observe the step number of testing.

3) Press

Shift

+ Trigger

and begin to operate, when the automatic testing is pause (the current prolong time of

one step is 25.5S), press

Shift + Trigger also can make the testing continue.

4)When the automatic testing is finished, the buzzer will hint by tweet (buzzer), VFD reads the result of testing. If the testing passed, VFD reads PASS, or reads FAULT. Press▲▼ also to observe the numerical value and result of every step.

5)Press ESC to back the automatic testing.

Setting the Resolution and number of decimal points

By selecting the low range and high range, you can measure the voltage and current more precisely.

34

Page 34
Image 34
B&K 8510 instruction manual Automatic Testing, Setting the Resolution and number of decimal points