Step 1

Press R-set

RESISTANCE =0000R

Step 2

enter a new value by using numeric keys or Rotary

RESISTANCE=*****R

 

knob to adjust the resistance value

 

Step 3

Press Enter to confirm.

0.000W R:0000R

V-set (set up a constant voltage from 1.5V to Max voltage)

8500A electronic load can be setup for a constant voltage. Constant voltage setup procedure is as following:

 

Procedure

 

Operation details

 

VFD display

 

Step 1

 

Press

V-set

 

VOLTAGE=1.500V

 

 

 

 

 

 

 

 

Step 2

 

enter a new value by using numeric keys or Rotary

 

VOLTAGE=*.***V

 

 

knob to adjust the resistance value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Step 3

 

Press Enter

to confirm.

 

0.000W

V:3.000V

Shift + Store

Procedure

Operation details

VFD display

 

 

 

 

 

Step 1

Press

Shift

and Store

STORE 1

Step 2

Press

Enter

to confirm.

Store the relative data

Shift + Recall

Procedure

Operation details

VFD display

 

 

 

 

 

Step 1

Press

Shift

and Recall

RECALL 1

Step 2

Press

Enter

to confirm

Recall the saving data

 

 

 

 

 

On/Off input setting

Use On/Off to change the state of electronic load. Switch on to off state by press On/Off .

Transition Testing Operation

Users could switch between the two different current and voltage level settings in the transition mode; it could test the transition specialty of power supply. Users could use front panel or communication interfaceTRAN ON AND TRAN OFFto make it work or not, please setting parameters before transition operation. Include Transition setting value, Constant pulse width setting and Transition Pulse width setting and Transition testing mode. The mix pulse width is 500uS.The Max pulse width is 6S. Transition Operation only could work in CC and CV mode.

Users can choose one of the three operation modes: Continuous, Pulse and Toggling mode.

28

Page 28
Image 28
B&K 8510 Transition Testing Operation, Set set up a constant voltage from 1.5V to Max voltage, Shift + Store