Von/Voff Operation

You can set voltage value Von/Voff to control the input state on/off for electronic load. When the input voltage reaches the Von value, the load’s input state is on. When the input voltage reaches the Voff value, the load’s input state is off.

Action for set Von/Voff value:

1Press Shift + Menu into menu

2VFD displays >CONFIG, presskey to choose >SYSTEM SET, press Enter to confirm

3VFD displays >MAX CURRENT SET, presskey to choose>VOLTAGE ON SET, press Enter key to

confirm

4VFD displays >VOLT.ON=0.00V, press numeric keys to set Von value (0.1V to max voltage value), press Enter

5Presskey to choose >VOLTAGE OFF SET, press Enter to confirm

6VFD displays >VOLT.OFF=0.00V, press numeric keys to set Voff value (0V to max voltage value), press Enter