8.4.5.6 CALibrate<chn>:PARameter:OFFSet<n>? [MIN|MAX|DEF]
Returns the calibration offset parameter for the specified channel.
CAL1:PAR:OFFS2?
–1.2
The CALibratesuffix, <chn>, specifies the channel (1 to 4). This command
does notapply to memory channels. For PRT/Thermistor channels, the calibra-
tion offsetsuffix <n> specifies the resistance range parameter for PRTs (1) or
Thermistors(2). ForThermocouple channels, the calibration offset suffix <n>
is ignored and may be omitted. Note:For PRT/Thermistor channels, the value
returned by this command is always in ohms.
ForPRT/Thermistor channels, specifying the MIN parameter returns a value of
-9 when the offsetsuffix <n> is 1 (for PRTs) and -9000 when the offset suffix
<n> is 2 (for Thermistors).For Thermocouple channels, specifying the MIN
parameter returnsa value of -9. For PRT/Thermistor channels, specifying the
MAX parameter returns a valueof 9 when the offset suffix <n> is 1 (for PRTs)
and 9000 when the offsetsuffix <n> is 2 (for Thermistors). For Thermocouple
channels, specifying the MAX parameter returns a valueof 9. Specifying the
DEF parameter returns a valueof 0 .
8.4.5.7 CALibrate<chn>:PARameter:OFFSet<n> <num>|MIN|MAX|DEF
Sets the calibration offset parameter for the specified channel.
CAL1:PAR:OFFS25
The CALibratesuffix, <chn>, specifies the channel (1 to 4). This command
does notapply to memory channels. For PRT/Thermistor channels, the calibra-
tion offsetsuffix <n> specifies the resistance range parameter (in ohms) for
PRTs(1) or Thermistors (2). For Thermocouple channels, the calibration offset
suffix<n> is ignored and may be omitted. The <num> paramter specifies the
valuefor the specified offset parameter. Specifying the MIN, MAX or DEF pa-
rameter setsthe offset parameter to the minimum, maximum, or default value
respectively.
This command is password-protected. Toaccess this command, the password
must besuccessfully entered u sing theSYST:PASS:CEN <password> com-
mand (see Section9.4.11, Password Commands).
8.4.5.8 CALibrate<chn>:PARameter:SCALe<n>? [MIN|MAX|DEF]
Returns the calibration scale parameter for the specifiedchannel.
CAL1:PAR:SCAL2?
0
The CALibratesuffix, <chn>, specifies the channel (1 to 4). This command
does notapply to memory channels. For PRT/Thermistor channels, the calibra-
tion scalesuffix <n> specifies the resistance range parameter for PRTs (1) or
101
8 Digital Communications Interface
Commands