day the measurement wasmade
Withthis format kohms and millivolts are not converted to ohms and volts.
This command may return the samereading if a new reading has not yet been
obtained.If only new readings are desired, the FORMat:STAMp command can
be used toenable the extended format option and any readings in which the
valueof the first field is 0 can be ignored or else the STATus:MEASure?com-
mand can be used to determine if a newreading is available for a desired
channel.
8.4.1.5 FORMat:STAMp?
Returnsthe extended measurement data format setting. See Section 9.4.1.4,
FETCh? for details.
FORM:STAM?
0
8.4.1.6 FORMat:STAMp<bool>
Enables extendedmeasurement data format with the FETCh?, MEASure? And
READ?commands. On powerup this is always set to off.
FORM:STAMON
8.4.1.7 MEASure? [<chn>]
Returns the most recent measurement for the specifiedchannel.
MEAS? 1
0.0127
This command is the sameas the FETCh? [<chn>] command
8.4.1.8 READ? [<chn>]
Returns the most recent measurement for the specifiedchannel.
READ? 2
0.0113
This command is the same as the?FETCh [<chn>] command.
8.4.1.9 SENSe<chn>:DATA?
Returnsthe input value for the specified channel.
SENS3:DATA?
100.0291, 0.0000
1529 Chub-E4 Thermometer Readout
User’s Guide
90