8.4.15 Conversion Typesand Parameters
Table7 lists the probe types, conversion types, mnemonics, and serial port re-
sponses. Table8 lists the conversioin types and their parameters.
ProbeType ConversionType Mnemonic Port Response
SPRT,PRT, RTD,
Thermistor
ITS-90(default)
ITS-SR5
PT-100
CVD
THRMR(T)
THRMT(R)
YSI-400
R(W)
ITS-90
ITS-SR5
PT100
CVD
THERM-R
THERM-T
YSI-400
RES
ITS
ITS5
PT
CVD
TRES
TTEM
YSI
RES
Thermocouple TC-K(default)
MilliVolts
TC-B
TC-E
TC-J
TC-N
TC-R
TC-S
TC-T
TC-AU/PT
Polynomial
TC-K
VIN[mV]
TC-B
TC-E
TC-J
TC-N
TC-R
TC-S
TC-T
TC-AU/PT
TC-POLY
K
V
B
E
J
N
R
S
T
AuPT
POLY
ConversionType Parameters
I90
ITS-SR5
PT100
CVD
YSI-400
RES
Range,Wires, RTPW, A4, B4, A, B, C, D
Range,Wires, RTPW, A5, B5
Range,Wires
Range,Wires, R0, AL, DE, BE
Range,Wires
RangeWires
THERM-R
THERM-T
B0,B1,B2,B3
A0,A1,A2,A3
1529 Chub-E4 Thermometer Readout
User’s Guide