By defaultthe password is set to ‘1529’ when the instrument ships from the
factory.We recommend that the user change the default password to protect the
password-protectedsettings. Note: Keep your password in a secure location
and do not forget it.
Toenter the password, press the Enter button. Press the LR buttons to high-
light the digit to change. Use the UD buttons to scroll through the numbers
0-9. Press the Enter buttonto save the new setting. Press the Exit button to
cancel and toexit to the menu.
If an incorrectpassword is entered, the display shows the message PASS-
WORD INCORRECT and access to thepassword-protected parameters is de-
nied. Press the Enter buttonto return to the menu. If the correct password is
entered, the bottom portion of the display allows the user to select the channel
to view.
Each channel of the1529 is calibrated independently and therefore each chan-
nel has its own set of calibration parameters which are displayed after the CAL
CH parameter isentered.
Use theLR buttons to select the channel. Press the Enter button to view the
calibration parameters for the selected channel. Press the Exit button to cancel
and to exitto the menu.
ForPRT/thermistor channels, the parameters C0, C100, C400, C0K, C10K,
C100K, DATE,and DUE are displayed.
The C0, C100, and C400 parameters apply to the PRTprobe types and
100Ωresistance range. Use these parameters to adjust the measurement at
0, 100 and 400Ωrespectively. These values are in ohms.
1529 Chub-E4 Thermometer Readout
User’s Guide
76
1
24.159C
3
24.142C
2
24.634C
4
24.015C
CAL CH: 1
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4THERMOMETER READOUT
Figure47 Select Calibration Channel