The <num> parameter is the number of measurements to be stored in the auto
log during the auto log session.Specifying the MIN, MAX, or DEF parameter
sets thenumber of measurements to auto log to the minimum, maximum, or de-
faultrespectively.
8.4.7.4 LOGging:AUTomatic:FREE?
Returnsthe number of auto log entries free and stored.
LOG:AUT:FREE?
8144,16
The response consistsof two numbers separated by a comma. The first number
indicates the number of auto log entries still available. The second number indi-
cates the number of entries already stored.
Note: Everytime an auto log sessio n isstarted , the firstentry in the log for that
session consistsof the data label and the date. All subsequent entries in the auto
log for that session consist of the channel, measurement value,units, and time.
Therefore, an auto log session that stores 100 measurements actually uses 101
log entries.
8.4.7.5 LOGging:AUTomatic:LABel? [MIN|MAX|DEF]
Returns the current auto log label setting.
LOG:AUT:LAB?
24
Specifying the MIN parameter returns a value of 1, Specifying the MAX pa-
rameter returnsa value of 25. Specifying the DEF parameter returns a value of
1.
8.4.7.6 LOGging:AUTomatic:LABel<num>|MIN|MAX|DEF
Sets the current auto log label.
LOG:AUT:LAB4
The <num>p arameterspecifies the auto log label number (1 to 25) to select.
Specifying theMIN, MAX, or DEF parameter sets the auto log label to the
minimum, maximum, or defaultrespectively.
8.4.7.7 LOGging:AUTomatic:POINt? [MAX]
Returnsthe number of auto log entries already stored.
LOG:AUT:POIN?
28
Specifying the MAX parameter returns the total number of entries that can be
stored inthe auto log.
107
8 Digital Communications Interface
Commands