the specifiedchannel. Specifying the MIN, MAX, or DEF parameter sets the
RJC parameter to the minimum, maximum, or default value respectively.
This command is password-protected. Toaccess this command, the password
must besuccessfully entered u sing theSYST:PASS:CEN <pass> command
(see Section9.4.11, Password Commands).
8.4.6 Display Commands
The display commands are used for querying and setting the display resolution,
lamp brightness,decimal format, window setting, and window fields.
8.4.6.1 DISPlay:RESolution? [MIN|MAX|DEF]
Returns the display resolution setting.
DISP:RES?
2
Specifying the MIN pamameter returns a value of 0. Specifying the MAX pa-
rameter returnsa value of 4. Specifying the DEF parameter returns AUT.
8.4.6.2 DISPlay:RESolution<num>|AUT|MIN|MAX|DEF
Sets the display resolution setting.
DISP:RES 3
The <num> parameter specifiesthe number of decimal places to display. This
parameter can be0, 1, 2, 3, or 4. Specifying theMIN, MAX or DEF parameter
sets thed isplayresolution to the minimum, maximum, or default value respec-
tively.Specifying the AUT parameter sets the display resolution to automatic.
8.4.6.3 DISPlay:LAMP? [MIN|MAX|DEF]
Returnsthe lamp brightness setting.
DISP:LAMP?
0
Specifying the MIN parameter returns a value of 0. Specifying the MAX pa-
rameter returnsa value of 3. Specifying the DEF parameter returns a value of 0.
8.4.6.4 DISPlay:LAMP <bool>|<num>|MIN|MAX|DEF
Sets thelamp brightness setting.
DISP:LAMPON
The lamp can beturned on or off by using either the <bool> parameter (ON or
OFF), the <num>parameter (0, 1, 2 or 3), or the MIN,MAX or DEF parame-
ters. Specifyingthe <bool> parameter to ON turns on the lamp. Specifying the
1529 Chub-E4 Thermometer Readout
User’s Guide
104