The <bool>parameter turns the keypad lockout on (1 or ON) or off (0 or OFF).
Specifying theMIN, MAX, or DEF parameter sets the keypad lockout to the
minimum, maximum, or defaultrespectively.
8.4.9.19 SYSTem:POWer:BATTery?
Returns the current battery charge level.
SYST:POW:BATT?
90
The returned valueindicates the current percentage of charge remaining in the
battery.The battery charge level is approximate and should be used only as a
guideline.
8.4.9.20 SYSTem:POWer:SOURce?
Returnsthe source of power being used.
SYST:POW:SOUR?
0
Indicateswhether the instrument is operating on the battery or the AC adapter.
This command returns a 1 if the AC adapter is connected or a 0 if the instru-
ment isrunning on the battery.The battery percentage is approximate and
should onlyb eused as a general guideline when determining the amount of
chargeremaining.
8.4.10 Date and Time CommandsThe date and time commands are used for querying and setting the instrument
date and timeand the format that is used when displaying the date and time.
8.4.10.1 DISPlay:DATE:FORMat? [MIN|MAX|DEF]
Returnsa number indicating the current date format.
DISP:DATE:FORM?
0
The returned valueis either a 0, 1, 2, or 3. See Table 11 on page 130, Date For-
mats, for a list of acceptable date formats. Specifying the MIN parameter re-
turns a value of 0. Specifying the MAX parameter returns a value of 3.
Specifying the DEF parameter returns a valueof 0.
8.4.10.2 DISPlay:DATE:FORMat <num>|MIN|MAX|DEF
Sets the date format setting.
DISP:DATE:FORM1
1529 Chub-E4 Thermometer Readout
User’s Guide
118