channel, the probe serialnumber and the conversiontype. The characterization
coefficientsand parameters are displayed after the conversion type is selected.
The PROBEparameter is for selecting any one of the four input channels:
CHAN1, CHAN2, CHAN3and CHAN4 or one of the ten probe memory chan-
nels: PMEM0, PMEM1, PMEM2, PMEM3, PMEM4, PMEM5, PMEM6,
PMEM7,PMEM8, and PMEM9. The probe memory channels are for storing
probe settingsthat can later be used to copy to an input channel.
The SERIAL parameter isthe serial number of the probe being edited. The se-
rialnumber consists of a string of up to eight characters using any numeric dig-
its,letters, and underscore.
The TYPE parameter is theconversion type for the corresponding input chan-
nel. The list of possible conversiontypes for each probe type is given in Table
2. The conversiontypes available are dependent on the probe or channel type as
indicated in the table. The characterization coefficientsand p arameters are dis-
played after the conversiontype is selected. These coefficients may change de-
pending on the selected conversiontype. The conversions and related
coefficientsand parameters are explained in the following sections.
The LR buttonsare used to select the parameter setting. The UD buttons
are used to scrollb etween and tochange each digit of the parameter. Press the
Enter buttonto change and save the new setting. Press and hold the Exit button
to cancel to the main display or press theEXIT button to cancel and to move to
37
7 Menu Functions
Probe Menu
1
24.159C
3
24.142C
2
24.634C
4
24.015C
URANGE:
100Ω
PROBE:
WIRES:
4
SERIAL:
145-560
RO:
100.0000
TYPE: CVD
DAL:
0.003850000
CHAN-1
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4THERMOMETER READOUT
Figure14 Edit Probe