hold theEx it buttonto cancel to the main display or press the EXIT button to
cancel and to moveto the next parameter.
7.1.3 Channel ModeThe CHANNELMODE function sets the channel display mode. When this
function isselected, the bottom portion of the display shows the current scan
mode and allows the setting to be changed.
The followingscan modes are available:
•SCAN - measures enabled channels scanning through and displaying the
channels one at a time, sequentially, in large digit format.
•SIMULT- measures and displays enabled channels simultaneously.
The LR buttonsare used to select a setting. Press theEnter buttontosave
the setting and exit. Press the Exit buttontocancelandtoexittothemenu.
If the measureperiod is 0.1 or 0.2 seconds, the CHANNEL MODE function is
unavailable.The channel mode is automatically set to scan mode. The follow-
ing message isdisplayed when the CHANNEL MODE function is selected.
NOTAVAILABLEAT THIS RATE.
Press ENTER to continue…
If the measure period is 0.5 seconds or greater,the channel mode operates
normally.
7.1.4 Moving AverageThe MOVINGAVERAGE function sets the input averaging.The moving aver-
age filteris u seful for smoothing variations in the measurements and for im-
33
7 Menu Functions
Channel Menu
1
24.159C
3
24.142C
2
24.634C
4
24.015C
SC MODE: SIMULT
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4THERMOMETER READOUT
Figure10 Channel Mode