sured error. For example, if the thermocouple temperature is actually
0.0°C but the 1529 reads 0.184°C, the CRJ parameter should be adjusted
by subtracting 0.184. If it was previously 0.000 it should now be –0.184.
6. Record the date with the DATE parameter. Set the DUE date to the de-
sired calibration interval.
7. Repeat the procedure for any other thermocouple channels.
1529 Chub-E4 Thermometer Readout
User’s Guide