The SENSe suffix,<chn>, specifies the channel (1 to 4). The parameter is
ohms for lowrange PRTs, Kohms for high range PRTs and thermistors, and
millivoltsfor thermocouples. The second returned parameter is always 0 for
PRTsand themistors and the reference junction temperature for thermocouples.
8.4.2 Measurement Control CommandsThe measurementcontrol commands are used to query and set the moving av-
erage filter,the resistance range, the number of probe wires, and the sequence
timer.The measurement control commands for resistance range and the number
of probe wires are passwordprotected.
8.4.2.1 INITiate
Provided for compatibility but has no effect.
INIT
8.4.2.2 INITiate:CONTinuous?
Provided for compatibility but has no effect.
INIT:CONT?
1
8.4.2.3 SENSe:AVERage:COUNt? [MIN|MAX|DEF]
Returns the moving averagefilter setting.
SENS:AVER:COUN?
4
Specifying the MIN parameter returns a value of 1, specifying the MAX pa-
rameter returnsa value of 10, and specifying the DEF parameter returns a value
of 1. The *RST command sets the averagecount to the default (1).
8.4.2.4 SENSe:AVERage:COUNt <num>|MIN|MAX|DEF
Sets the moving averagefilter setting.
SENS:AVER:COUNMIN
The <num>parameter must be a number between 1 and 10. Specifying the
MIN, MAX, or DEF parameter sets the moving average filtersetting to the
minimum, maximum, or default,resp ectively.The *RST command sets the
movingaverage to 1.
8.4.2.5 SENSe<chn>:RESistance:RANGe? [MIN|MAX|DEF]
Returnsthe resistance range setting.
SENS3:RES:RANG?
91
8 Digital Communications Interface
Commands