The C0K, C10K, and C100K parameters apply to the thermistor probe
types and 10KΩresistance range. Use these parameters to adjust the mea-
surement at 0K,10K, and 100KΩrespectively. These values are inKohms.
The DATE parameter is the date the current calibration was performed.
The DUE parameter is the date the current calibration expires.
Forthermocouple channels, the parameters C0, C100, CRJ, DATE, and DUE
are displayed.
The C0 and C100 parameters adjust the measurement at 0 and 100 mV re-
spectively. These values are in mV.
The CRJ parameter adjusts the reference junction temperature measure-
ment. This parameter is in °C.
The DATE parameter is the date the current calibration was performed.
The DUE parameter is the date the current calibration expires.
Press the LR buttons to change a setting and the UD buttons to scroll be-
tween parameters. Press the Enter buttonto save the new setting. Press the
Exit buttonto exit to the menu.
7.5.5 System Reset
The SYSTEM RESETfunction r esetsall operating parameters to the factory
defaultvalues. This function is password-protected (see Section 7.5.3, Pass-
word).In order to access the system reset, you must enter the correct four-digit
password.When this function is selected, the bottom portion of the display
prompts the user to enter the current password togain access to this option.
77
7 Menu Functions
System Menu
1
24.159C
3
24.142C
2
24.634C
4
24.015C
ENTER CORRECT PASSWORD FOR ACCESS.
PASSWRD: 0
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4THERMOMETER READOUT
Figure48 Pass Cal Reset