Press Clear (F4) to erase all or selected test results.

Press the ESCAPE key to return to the EXTENDED FUNCTIONS screen. See Chapter 3: Autotest for more information.

Autotest Setup

To display the AUTOTEST SETUP screen, press the Extended Functions key, then select Setup Autotest. This screen lets you select the default Autotest, edit Autotest settings and delete user- defined Autotests.

Press the or arrow key to scroll through the list of Autotests (or to page up and to page down). When the Autotest you want is highlighted, press Default (F1) to select it as the default Autotest.

Note: An Autotest name proceeded by an asterisk (*) indicates that the original Autotest has been modified and the Autotest name has not been changed.

Press Edit (F2) to change the Autotest name, to select pairs to be used, to select the tests to run, and to set test limits.

Press Delete (F3) to delete the highlighted user-defined Autotest.

Press the ESCAPE key to return to the SETUP screen.

See Chapter 3: Autotest for more information.

Editing Autotest Information

To display the EDIT AUTOTEST screen, press the Extended Functions key, then select Setup Autotest, then press Edit (F2). This screen lets you customize Autotest settings.

Press the or arrow key to scroll through the available options and press Select (F1) to choose the highlighted option. Or, press the number on the keypad that corresponds to the option you want.

Select Name/Options to edit the Autotest name and to change the Margin Options.

Select Pairs to choose the pairs to be tested during Autotest, and to indicate which pairs have PASS/FAIL criteria applied to them during the Autotest.

Select Tests to Run to choose the tests to run during Autotest.

Select Limits to edit test limits.

Press the ESCAPE key to return to the AUTOTEST SETUP screen.

Calibrating the Injector

The Injector must be calibrated each calendar day that you use it to ensure accurate measurements. To calibrate the Injector, connect the Scanner and the Injector with the patch cable supplied with the Scanner. Press the Extended Functions key; then select Calibrate Injector.

Press Cancel (F1) or the ESCAPE key to stop the calibration.

See Chapter 2: Getting Started for more information on the Injector.

Page 48
Image 48
Microtest TSB-67 - 568A Annex E manual Autotest Setup, Calibrating the Injector