Contrast Medium
131
Workflow
To start a contrast enhanced examination in coupled
mode:
• Select the Scan subtask card.
• Select under the menu field Scan Start either the
entry Injector coupled (Start button) or the entry
Injector coupled (Footswitch)
– Injector coupled (Start button): The Start button
of the CT scanner will start the injector.
– Injector coupled (Footswitch): The footswitch of
the CT scanner will start the injector.
If an injector is connected, load the scan mode first and
then arm the injector.
Depending on the injector it might be not possible to
arm the injector before the scan protocol is loaded (see
User manual of the injector). When the mode is
loaded, the CT scanner will ask you to check the injec-
tor and to arm it. Check the parameters at the injector
side and confirm the parameters. The injector is armed
and ready for the examination.
Note: If the scan mode is unloaded the injector will
also be disabled.