Vascular
415
Hints
• CARE Bolus may be used to optimize the bolus tim-
ing.
• Set the ROI for monitoring scan with triggering
threshold of 120 HU, or use manual triggering.
• If Topo length 1024 mm is not long enough, you can
also choose the 1540 mm long Topogram.
• Position the patient as feet first. Bend the feet
together if necessary.
Contrast medium IV injection
Start delay 10 – 20sec.
Flow rate 3.0 – 3.5 ml/sec.
Total amount 120 – 150 ml