500
Respiratory Gating
Scan ProtocolRespSeq
Indications:
A sequential mode for studies with respiration trigger-
ing.
For SOMATOM Emotion 16-slice configuration:
The scan length is predefined with 30.5 cm.
For SOMATOM Emotion 6-slice configuration:
The scan length is predefined with 30.3 cm.
Emotion 16 RespSeq
kV 130
Effective mAs/
Quality ref. mAs
16
Rotation time 0.6sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2mm
Slice width 2.4 mm
Feed/Scan 19.2mm
Kernel B31s
CTDIvol 1.79mGy
Effective dose Male: 0.92mSv
Female: 1.18 mSv