Head
199
Hint
• For low dose studies use only 20 mAs.
Emotion 6 Orbit 2nd reconstr.
kV 130
Effective mAs/
Quality ref. mAs
70
Rotation Time 1 .0 sec.
Acquisition 6 x 1.0mm
Slice collimation 1.0 mm
Slice width 5.0 mm 1.25mm
Feed/Rotation 5.1 mm
Pitch Factor 0.85
Increment 5.0mm 0.5 mm
Kernel H70s H70s
CTDIVol 19.25mGy
Effective dose Male: 0.25 mSv
Female: 0.32mSv
Contrast medium IV injection
Start delay 45 sec.
Flow rate 2.0 ml/sec.
Total am ount 75 m l