568
Children
ExtrRoutineHR
Indications:
Spiral mode for high-resolution bone studies, for
example, tumors, post-traumatic changes, orthopedic
indications, etc.
For SOMATOM Emotion 16-slice configuration:
A typical range of 6 cm covered in 35.33 sec.
For SOMATOM Emotion 6-slice configuration:
A typical range of 10 cm covered in 35.33 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 60 mm was used.
** Adjust the mAs value to the body region.
Emotion 16 ExtrHR
kV 110
Effective mAs/
Quality ref. mAs**
50
Rotation time 1.0 sec.
Acquisition 4 x 0.6 mm
Slice collimation 0.6mm
Slice width 0.6 mm
Feed/Rotation 1.8mm
Pitch Factor 0.75
Increment 0.4 mm
Kernel U90 s
CTDIVol 4.70 mGy
Effective dose Male: 0.01mSv*
Female: 0.01mSv*