198
Head
Orbit
Indications:
Spiral mode for orbital studies, e.g. fracture.
For SOMATOM Emotion 16-slice configuration:
A range of 50 mm will be covered in 8.51sec.
For SOMATOM Emotion 6-slice configuration:
A range of 50 mm will be covered in 11.80sec.
Emotion 16 Orbit 2nd rec onstr.
kV 130
Effective mAs/
Quality ref. mAs
70
Rotation Time 1.0 sec.
Acquisition 16 x 0.6mm
Slice collimation 0.6 mm
Slice width 5.0 mm 1.0mm
Feed/Rotation 7.7mm
Pitch Factor 0.80
Increment 5.0mm 0.7m m
Kernel H70s H70s
CTDIVol 18.62mGy
Effective dose Male: 0.50 mSv
Female: 0.54mSv